Table 14. Troubleshooting
Problem Possible Cause Solution
Detector Operation
Detector does not display expected
gas readings after activation self-test.
Sensor not stabilized Used sensor: wait 60 seconds
New sensor: wait 5 minutes
Sensor(s) requires calibration Calibrate the sensor(s). Refer to Calibration
.
Target gas is present Detector is operating properly. Use caution in
suspect areas.
Detector does not respond to
pushbutton.
Battery is depleted Charge battery. Refer to Charging the Battery
.
Detector is performing operations that do
not require user input
Pushbutton operation restores automatically
when the operation ends.
Detector does not accurately measure
gas.
Sensor(s) requires calibration Calibrate the sensors. Refer to Calibration
.
Detector is colder/hotter than ambient gas Allow the detector to attain ambient tempera-
ture before use.
Sensor filter is blocked Replace the sensor filter. Refer to Replacing a
Sensor or Sensor Filter.
Detector does not enter into alarm. Alarm setpoint(s) defined incorrectly Reset alarm setpoints. Refer to Factory Gas
Alarm Setpoints and Sensor Configuration.
Alarm setpoint(s) set to zero Reset alarm setpoints. Refer to Factory Gas
Alarm Setpoints and Sensor Configuration.
Detector is in calibration mode Complete the calibration procedure.