BW Technologies MicroClip XT Dokumentacja Strona 26

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 25
15
GasAlertMicroClip XT
Activating the Detector
Last Bump Test Failed
If any sensor failed the last calibration, CAL FAILURE displays
on the screen.
Note
When the BUMPCHK FAILURE displays, the previous calibra-
tion has failed but the calibration is still valid until the next
bump test due date.
Force Bump (optional)
8. If the Force Bump option is enabled and the sensors are
due have a bump test, the following screen displays.
A bump test must be performed to enter normal operation. Apply
gas to the sensors. Ensure the visual, audible, and vibrator
alarms activate. When the gas is removed, the detector briefly
remains in alarm until the gas has cleared from the sensors.
When the sensors successfully pass the bump check, the follow-
ing screen displays showing the number of days remaining until
the next bump check is due (I d = 1 day).
If Force Bump is disabled, press C to continue with the startup
self-tests.
Note
If BUMPCHK todAY displays again after performing a bump
check, refer to Startup Troubleshooting
.
Przeglądanie stron 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 73 74

Komentarze do niniejszej Instrukcji

Brak uwag