BW Technologies MicroClip XT Instrukcja Użytkownika

Przeglądaj online lub pobierz Instrukcja Użytkownika dla Systemy kontroli dostępu bezpieczeństwa BW Technologies MicroClip XT. BW Technologies MicroClip XT User manual Instrukcja obsługi

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj

Podsumowanie treści

Strona 1 - GasAlertMicroClip

User ManualGasAlertMicroClip1, 2, 3, and 4 Gas Detectors D5908/2 EnglishiERP: 124291© BW Technologies 2007, all rights reserved. Printed in Canada.A

Strona 2

GasAlertMicroClip User Manual Contacting BW Technologies by Honeywell To contact BW Technologies by Honeywell, call: USA: 1-888-749-8878 Canada: 1-800

Strona 3 - Table of Contents

GasAlertMicroClip Safety Information - Read First aCautions⇒ Warning: Substitution of components may impair Intrinsic Safety. ⇒ Caution: For saf

Strona 4 - Title Page

GasAlertMicroClip User Manual aCautions⇒ Protect the combustible sensor from exposure to lead compounds, silicones, and chlorinated hydrocarbons. Alt

Strona 5 - List of Tables

GasAlertMicroClip Safety Information - Read First Table 2. International Symbols Symbol Meaning nApproved to both U.S. and Canadian Standards by C

Strona 6

GasAlertMicroClip User Manual Getting Started The list below provides the standard items included with your detector. If the detector is damaged or so

Strona 7 - List of Figures

GasAlertMicroClip Getting Started Figure 1. Parts of the GasAlertMicroClip Table 3. Parts of the GasAlertMicroClip Item Description 1 Visual ala

Strona 8

GasAlertMicroClip User Manual Table 4. Display Elements Item Description 1 Alarm condition 2 Automatically span sensor 3 Gas cylinder 4 Gas identifie

Strona 9

GasAlertMicroClip Getting Started Table 5. Pushbutton Pushbutton Description C• To activate the detector press C. • To deactivate the detector,

Strona 10 - Honeywell

GasAlertMicroClip User Manual Activating the Detector To activate the detector, press C in a normal clean atmosphere. Self-Test Once the detector is a

Strona 11

GasAlertMicroClip Activating the Detector Note Users can enter a 2-line message, up to a maximum of 25 characters per line. The messag

Strona 12

Limited Warranty & Limitation of Liability BW Technologies LP (BW) warrants this product to be free from defects in material and workmanship under

Strona 13 - Symbol Meaning

GasAlertMicroClip User Manual Note The sensors are continuously tested while the detector is activated. Automatic Zero and O2 Calibration (optional) 5

Strona 14 - Getting Started

GasAlertMicroClip Activating the Detector Press C to acknowledge the warning and enter normal operation, or continue with the self-test if appli

Strona 15 - Item Description

GasAlertMicroClip User Manual Bump Check (optional) Note Bump check is not performed if the detector was just calibrated (step #6). BW recommends to “

Strona 16 - Figure 2. Display Elements

GasAlertMicroClip Activating the Detector Self-Test Fail While the LCD displays BUMPCHK todAY, apply gas to the detector or press C to deactivate

Strona 17 - Pushbutton Description

GasAlertMicroClip User Manual Backlight The LCD’s backlight automatically activates when there is an alarm condition. Pressing C activates the backlig

Strona 18 - Activating the Detector

GasAlertMicroClip Options Sensor Disabled Detector Identification a Warning Start-up Message Disabling an installed sensor configures the detector

Strona 19

GasAlertMicroClip User Manual Bump Interval Calibration Gas Concentration The Bump Interval determines the number of days before a sensor needs to be

Strona 20

GasAlertMicroClip Options Automatic O2 Calibration TWA Alarm When the O2 Auto-Calibration on Startup option is enabled, it forces the detector to a

Strona 21

GasAlertMicroClip User Manual User Options Menu Confidence Beep The Confidence Beep notifies that the detector is on and the battery has sufficient po

Strona 22 - Bump Check (optional)

GasAlertMicroClip Options Low Alarm Acknowledge Force Bump When Overdue When Low Alarm Acknowledge is enabled, the audible alarm can be disabled du

Strona 23

Table of Contents Title Page Introduction...

Strona 24 - Options

GasAlertMicroClip User Manual If more than one type or level of alarm exists at the same time, a multi-gas alarm will result. Alarms The following tab

Strona 25 - Sensor Options

GasAlertMicroClip Alarms Table 6. Alarms Alarms Display Alarms Display Multi-Gas Alarm: • Alternating low and high alarm siren and flash • L and t

Strona 26

GasAlertMicroClip User Manual Table 6. Alarms Alarms Display Normal Shutdown: • Four beeps and flashes • Vibrator alarm temporarily activates Note

Strona 27

GasAlertMicroClip Alarms Viewing Gas Exposures To view the TWA, STEL, and maximum (MAX) hold readings, press C twice. The LCD first displays the TW

Strona 28

GasAlertMicroClip User Manual Gas Alarm Setpoints The detector's gas alarm setpoints trigger the gas alarms that are described in the table below

Strona 29

GasAlertMicroClip Alarms Stopping a Gas Alarm The low and high alarms stop when the ambient gas level returns to the acceptable range. Note If alar

Strona 30 - Table 6. Alarms

GasAlertMicroClip User Manual Calibration • Calibrate the detector at least once every 180 days, depending on use and sensor exposure to poisons and

Strona 31

GasAlertMicroClip Calibration Wind currents may cause false readings and poor calibrations. Diagnostics Protection The detector tests the ambient a

Strona 32 - Computed Gas Exposures

GasAlertMicroClip User Manual Calibration Procedure To calibrate the detector perform the following procedure. Note Calibrate O2 in clean air. If C is

Strona 33

GasAlertMicroClip Calibration Auto Zero and Oxygen Sensor Calibration Auto Span 4. The LCD now prompts you to apply calibration gas to the sensors

Strona 34 - Gas Alarm Setpoints

GasAlertMicroClip User Manual ii Title PageLow Battery Alarm ...

Strona 35 - Automatic Shutdown Alarm

GasAlertMicroClip User Manual No Gas Detected When the detector senses a sufficient amount of gas concentration (approximately 30 seconds), the audibl

Strona 36 - Calibration

GasAlertMicroClip Calibration If you apply gas to a sensor and the detector fails to span the sensor, repeat the calibration procedure using a new

Strona 37 - Applying Gas to the Sensors

GasAlertMicroClip User Manual Verification • Peak exposure level (ppm or %); After calibration is complete and the detector is in normal operating mo

Strona 38

GasAlertMicroClip Maintenance Charging the Detector a Warning ⇒ The detector must be charged in a non-hazardous atmosphere of 32°F - 113°F (0°C -

Strona 39 - Auto Zero Sensor Fail

GasAlertMicroClip User Manual Each sensor has a high degree of resistance to common vapors and gases. A sensor will most likely clear itself if you mo

Strona 40

GasAlertMicroClip Maintenance 3. Remove the two screws on the PCB and remove the PCB. Table 11. Replacing a Sensor or Sensor Filter Item Descripti

Strona 41

GasAlertMicroClip User Manual 6. Re-assemble the detector. When assembling the detector, be aware of the following: • Ensure the charging contact pi

Strona 42 - Event Logs

GasAlertMicroClip Troubleshooting Troubleshooting If you are still unable to correct the problem, contact BW Technologies by Honeywell. The detect

Strona 43 - Charging the Detector

GasAlertMicroClip User Manual Table 12. Troubleshooting Tips (cont.) Problem Possible Cause Solution Detector does not display normal ambient gas read

Strona 44

GasAlertMicroClip Troubleshooting Table 12. Troubleshooting Tips (cont.) Problem Possible Cause Solution Detector does not enter alarm. → Alarm se

Strona 45

List of Tables Table Title Page 1. Gases Monitored ...

Strona 46

GasAlertMicroClip User Manual Replacement Parts and Accessories a Warning To avoid personal injury or damage to the detector, use only the specified r

Strona 47 - Troubleshooting

GasAlertMicroClip Specifications Model No. Description Qty MC-VC-K1 Vehicle charging kit 1 MC-C01-MC5*GasAlertMicroClip MultiCharger 1 GA-USB1-

Strona 48

GasAlertMicroClip User Manual Approvals: Approved by CSA to both U.S. and Canadian Standards CAN/CSA C22.2 No. 157 and C22.2 152 ANS/UL – 913 and ANSI

Strona 49

GasAlertMicroClip Specifications General Datalogger Specifications Storage: 16 hours at 15-second intervals Memory type: Wraparound memory ensures

Strona 50 - Model No. Description Qty

GasAlertMicroClip User Manual 46

Strona 51 - : 10.75 x 6.00 x 2.73 cm

User ManualGasAlertMicroClip1, 2, 3, and 4 Gas Detectors D5908/2 EnglishiERP: 124291© BW Technologies 2007, all rights reserved. Printed in Canada.A

Strona 52

Printing InformationiERP # 125109Size: 3¾ x 5½ inchesStock: 80 lb cougar cover card stockPrint Color: BlackLanguage: English

Strona 53

DAILY BUMP TEST iERP: 125538BW recommends to “bump test” the sensor(s), before each day’s use to confi rm their ability to respond to gas by expos-in

Strona 54 - User Manual

GasAlertMicroClip User Manual iv

Strona 55

List of Figures Figure Title Page 1. GasAlertMicroClip Detector...

Strona 56 - ERRATA CARD

GasAlertMicroClip User Manual vi CAUTION: FOR SAFETY REASONS, THIS EQUIPMENT MUST BE OPERATED AND SERVICED BY QUALIFIED PERSONNEL ONLY. READ AND UND

Strona 57 - DAILY BUMP TEST

GasAlertMicroClip Introduction a Warning To ensure your personal safety, read “Safety Information” before you use the detector. The GasAlertMicroCli

Komentarze do niniejszej Instrukcji

Brak uwag